Model |
STA-1225 |
STA-1150 |
Max. loading capacity |
2.5kN (250kgf) |
500N (50kgf) |
Effective stroke |
850mm |
550mm |
Crosshead speed (mm/min.) |
0.5~1000 (14 speeds) |
0.5~500 (13 speeds) |
Crosshead random speed |
Set with 0.1 mm pitch. Up to 6 settings can be stored. |
Crosshead speed accuracy |
+/-0.1% |
Return speed |
1000, 500 |
500, 100 |
Crosshead manual operation |
Position the crosshead first using UP / Down switches & use manual button to adjust fine position. |
Dimensions
(W x H x Dmm) |
470 x 1450 x 335 |
470 x 1050 x 335 |
Weight |
55kg |
44kg |
Moving Crosshead Position Display (mm) |
5 digital display (resolution : 0.01) |
Elongation Display used with Extenso-meter |
5 digital display (Possible to display Extension), Unit: mm, %GL, %FS |
Measurement Accuracy (100~10%) |
+/- 1% of indicated value or 0.2% of full scale whichever is greater |
Load Range |
1, 2, 4, 10, 20, 40, 100% (7 steps) provided with auto-ranging function |
Load Calibration |
1) Auto-zeroing, 2) Auto-spanning, 3) Load cell discrimination |
Over Load Safety Function |
Provided |
Sample Break Detection |
Provided |
Stroke Safety Device |
Upper/Lower limiters |
Cycle Test |
Upper/Lower limit points: select from either the crosshead position or load
Set momentary stop time at each upper/lower limit.
|
Power Source |
AC100V 1P 50/60Hz with 2P plug |